Cooper-instruments-systems DFI INFINITY Digital Force Indicator/Controller User Manual Page 78

  • Download
  • Add to my manuals
  • Print
  • Page
    / 78
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 77
CF 125 INFINITY SC GUIDE 74 M1519/N/0605
Table 11.34. Block E For “P” and “W” Prefix Commands For the Batch meter, Square-root and Rate
Modes
ITEM (HEX)
SUFFIX
DESCRIPTION
BYTE NO. &
ORDER
32 Calibration current span 1, 2
31 Calibration current zero 3, 4
30 Calibration volt span 5, 6
2F Calibration volt zero 7, 8
2E Calibration Oscillator Frequency 9
11.32.7 Example
EXAMPLE: Assume you want to send raw values for the batch meter #15 hex as below:
Batch value: 785,424=BFC10 Hex
# of Batches: 754=2F2 Hex
Total value: 592,209,696=234C6720 Hex
Computer sends: “*P420002F2000BCF10234C6720<CR>”
If you are in echo mode, you will receive:
“P42<CR>”